Indian Journal of Pharmaceutical Education and Research logo
Article

Design, Development, Impurity Characterization for Comprehensive Impurity Profiling in Eprosartan Mesylate by a Robust LC-MS/MS-Q-TOF Method: Emphasis on EPR-I, EPR-II, and Regio-Isomers

1, 2, 3★★ Corresponding author

  1. 1Department of Pharmaceutical Analysis, SVU College of Pharmaceutical Sciences, Sri Venkateswara University, Tirupati, Andhra Pradesh, INDIA.
  2. 2Department of Pharmaceutical Analysis, TRR College of Pharmacy, Meerpet, Hyderabad, Telangana, INDIA.
  3. 3Department of Pharmaceutical Analysis, Shri Vishnu College of Pharmacy (Autonomous) Affiliated to Andhra University, Bhimavaram, Andhra Pradesh, INDIA.

CORRESPONDENCE

Atla Srinivasa Rao

Department of Pharmaceutical Analysis, Shri Vishnu College of Pharmacy (Autonomous) Affiliated to Andhra University, Bhimavaram, Andhra Pradesh, INDIA.

vasanthpharma@gmail.com

Received: 03-07-2026; Revised: 21-08-2026; Accepted: 16-10-2026.

Volume 61, Issue 1 · pp. 331-338 · 2026 · DOI: 10.5530/ijper.20273045

ABSTRACT

Background: Eprosartan mesylate along side its related impurities, EPR-I, EPR-II, and their regioisomers, were simultaneously identified and quantified utilizing a sensitive and selective LC-MS/MS technique in conjunction with high-resolution mass spectrometry employing a Q-TOF analyzer.Materials and Methods: A reversed-phase chromatographic separation was made on Hypersil BDS Cl8, (250 x 4.6 mm), 5 µm at first, the composition of the mobile portion. A was 75% and B was 25%. Over 5 min, it was maintained constant. Subsequently, a linear gradient was applied from 5 to 19 min, increasing the proportion of B to 55%. Later was then held constant for 1 min before returning to the initial conditions of 75% A and 25% B, using ammonium acetate buffer and an acetonitrile with run time of 25 min, maintaining sampler cooler temperature at 5ºC. Electrospray ionization (+ve) was deployed. Other source parameters were optimized as nebulizer gas pressure at 13 psi, curtain gas pressure at 12 psi, ion spray voltage at 4500 V, temperature at 325ºC, focusing potential at 260 V, declustering potential at 55 V, and entrance potential at 10 V. A Q1 Multiple Scan type was used for data acquisition.Results: Analysis revealed the presence of Eprosartan at 3.12 min with an m/z of 425, EPR-I and its Regio isomer had similar molecular weights (m/z 335) but different retention times (16.56 and 12.81 min, respectively). Similarly, EPR-II and its Regio isomer shared a molecular weight (m/z 301 and 305, respectively) but had distinct retention times (11.88 and 11.03 min). As confirmed in compliance with ICH requirements, exhibiting exceptional specificity, EPR-I linear from LOQ-150% (r2 =0.9988), EPR-II linear from the LOQ-150% (r2 =0.9985), EPR-I regioisomer linear from the LOQ-150% (r2 =0.9985) demonstrated to be satisfactory by percentage recoveries at the LOQ level that fall between 98.0 and 103.0. The impurities had LOD and LOQ values of 0.06-0.1 ppm as well as 0.02-0.03 ppm, in each case.Conclusion: Aforesaid approach was successfully used to analyze Eprosartan Mesylate medicinal material and pharmaceutical formulations.

KEYWORDS

0% READ

FULL TEXT

INTRODUCTION

The growing rigor with which regulatory agencies monitor and manage the quality and purity of pharmaceuticals has piqued the interest of researchers all over the world in impurity profiling.1,2 and degradation studies of medicinal compounds.3–5 and pharmaceutical products. In addition, generic medications and products are now required by law to undergo impurity profiling and a forced or stress degradation research. Two crucial considerations in pharmacological treatment are the medicine's efficacy and safety.6,7 Impurity and DP qualifying thresholds, reporting, and detection are covered in ICH guidelines Q3A and Q3B. One medication that can help lower blood pressure is Eprosartan mesylate, an ARB. Traditional methods have a few drawbacks, such as the fact that they are laborious and can become more so if numerous pollutants and DPs need to be detected in a single sample. The procedure becomes more time-consuming if the impurity or DPs produced are in minute quantities that cannot be enhanced. These factors have made hyphenated approaches the go-to option for identification and characterization in recent years.8–12 particularly when impurities or DPs are present at trace levels. Eprosartan,13–20 is commonly administered alone or in conjunction with other anti-hypertensive medicines. While several UV,21–25 HPLC,26–33 methods have been developed for the analysis of Eprosartan-related impurities, and its relevant molecules in their tablet formulations, in human plasma and urine, these methods often rely on UV detection, which may not be sufficiently sensitive or specific for the detection of trace impurities. LC-MS/MS.34–47 offered superior sensitivity, selectivity, and specificity compared to traditional HPLC methods but couldn't clarify on sensitive detection of EPR-I, EPR-II, and their Regio isomers. By coupling LC-MS/MS with high-resolution mass spectrometry using a MS-Q-TOF analyzer, it is possible to detect and identify a wide range of impurities, including unknown impurities. The specific focus on EPR-I, EPR-II, and their Regio isomers is important as these have significant role on pharmacological and toxicological implications. It is feasible to guarantee that the pharmaceutical product satisfies strict quality standards by creating a sensitive and precise technique for the detection and measurement of these contaminants.

MATERIALS AND METHODS

Materials

EPR-I, EPR-II, EPR-I Regio isomer, EPR-II Regio isomer, and Eprosartan were sourced from Zydus Cadila Health Care Ltd in Secunderabad, India. LCMS-grade methanol, acetonitrile, and other required chemicals were supplied by the Merck Chemical Division in Mumbai. HPLC-grade water, purified using a Milli-Q water system, was used throughout the experiment. The chromatographic study took place out utilizing a Waters 2695 HPLC system, which was connected to a SCIEX QTRAP 5500 Mass Spectrometer and comprised of a high-speed autosampler, column oven, and degasser. Empower-2 software was utilized for data acquisition and analysis.

Methods

The following standards were carefully measured and put into individual 100 mL volumetric flasks EPR-I, EPR-II, EPR-I Regio isomer, and EPR-II Regio isomer. Totaling around 10 mg for each standard.

A main solution with a concentration of 100 mg/L (or 100 ppm) was prepared by dissolving the standards and diluting them with the diluent until they reached the mark.

First Dilution: A 100 mL volumetric flask was used to dilute the primary solution to the mark after 1.0 mL was pipetted into it. This gave a concentration of 1 mg/L (or 1 ppm).

To prepare the sample, about 300 mg equivalent was carefully weighed and transferred to a 10 mL volumetric flask. Diluting the sample to volume was the next step after dissolving it.

Primary Solution Preparation: Weighing and Dissolving:

Dilution to Volume:

Secondary Dilution:

Evaluation of System

A blank (diluent), benchmark preparation, and sample setup six times into the chromatograph, recorded the chromatograms of each impurity, and computed the content using the following formula:

AT: Area counts of EPR-I peak.

AS: Average counts of EPR-I peak.

CS: Concentration of EPR-I.

CT: Concentration of sample.

P: The degree of purity of the EPR-I product.

Specificity

Eprosartan Mesylate and all impurity solutions were prepared independently in the diluent at a concentration of approximately 0.1 mg/mL during the solution preparation process.

LOD and LOQ

The Limit of Detection (LOD) and Limit of Quantification (LOQ) were determined by comparing the results from blank samples with those from samples with known analyte concentrations and calculating the signal-to-noise ratio. This allowed us to identify the lowest level (10 μg/mL) at which the analyte could be reliably identified.

Precision at LOQ

Six replicate influxes of a typical solution were examined in order to assess the accuracy at the LOQ level.

Linearity

By examining the solutions ranging from LOQ to 150% of the specified limit, the LC-MS/MS method's linearity was verified.

Accuracy

By employing the diluent, 10 mL of the 1 ppm solution was subsequently reduced to a total volume of 100 mL to achieve the desired concentration of 0.1 ppm. Intermediate Dilution: To reach a concentration of 0.06 ppm, 6 mL of the 1 ppm solution was diluted with 100 mL of the diluent.

RESULTS

Optimization of LC as well as Mass Situation

A Hypersil BDS C18 column (250 x 4.6 mm x 5 µm) was utilized. The material was detected with a UV detector set to 220 nm. The mobile phase ran through the column at 1.0 mL/min, and the column oven was kept at 35ºC to achieve proper separation and peak form. To prevent degradation, samples were injected at 30 µL and stored at 5ºC in a sample cooler. The overall run time for the analysis was 25 min, and methanol was utilized as the diluent to effectively dissolve the materials. For the mass spectrometric analysis of Eprosartan and its impurities, utilizing an electrospray ionization (+ve). The nebulizer gas was set to a pressure of 13 psi, providing the necessary pressure to nebulize the sample into fine droplets. A curtain gas pressure of 12 psi was employed to prevent contamination of the ion source. Maintaining the ion spray voltage at 4500 volts is essential for the analyte's ionization. To facilitate the desolvation of the analyte ions, the ionization source's temperature was preset at 325ºC. In order to guarantee that the ions were effectively focused into the mass analyzer, the focusing potential was adjusted to 260 volts. Before the ion stream entered the mass spectrometer, solvent molecules and any other neutral species were eliminated using a declustering potential tuned at 55 volts. Ions entered the mass analyzer more easily because the entrance potential fostered at 10 volts. The Q1 Multiple (Q1) scan type was used, which allowed for the selective identification and measurement of the relevant ions.

System suitability

Six replicate injections of a standard solution were evaluated in order to determine the system's appropriateness. The peak location for Eprosartan Mesylate impurities appears in Table 1.

Specificity

The Retention Time (RT) values and Mass Number (m/z) values as displayed in Table 2 were obtained for Eprosartan and impurities representative selective ion recorded chromatograms as shown in Figure1.

LOD and LOQ

The minimum level at which the analyte can be consistently recognized is established by assessing the signal-to-noise ratio, achieved through a comparison of results from samples with known analyte concentrations against those of blank samples. A tripartite solution was devised and examined to determine the quantitation threshold in relation to the detection limit. Table 1 presents the results that were achieved.

Precision at LOQ

Table 1 summarizes the peak area of Eprosartan Mesylate impurities.

Linearity

Results obtained are shown in Table 2 shown the line of best fit for peak area versus concentration for Eprosartan Mesylate impurities.

Accuracy at LOQ

By evaluating Eprosartan Mesylate spiked with its impurities at the LOQ level, the accuracy at LOQ was ascertained. Each solution has been created in three instances and examined separately. Table 2 reveals the percent of recuperation findings for the impurities.

Accuracy at specification level

By evaluating Eprosartan Mesylate spiked with its impurities at the specification level, the correctness was ascertained. Each solution was made in triple and examined separately. In Table 2 the proportion of recuperation results for impurities are exemplified. 2

Table 1 Summary of Analytical Performance of Eprosartan and Its Isomers.
CompoundMean Peak Area (±SD)rLOD (μg/ mL)LOQ (μg/mL)S/N (LOD)S/N (LOQ)% Recovery (LOQ)% RSD
EPR I306,983±6,6772.180.030.103.5:110.4:1100.02.00
EPR II774,283±10,2891.330.030.103.1:19.8:1100.31.52
EPR Regio501,950±11,4242.280.020.063.3:110.1:1100.01.67
EPR II Regio3,048,667±97,8523.210.030.103.5:19.9:1102.00.98
Table 2 Chromatographic and Linearity Data of Eprosartan and Its Isomers.
CompoundRT (min)m/zConc. (µg/mL)Peak Area% Recovery (Spec.)%RSD (Spec.)
EPR I16.563350.1-3.015,120- 472,10098.4 - 101.92.62
EPR II11.883010.1- 3.042,970- 1,261,000Not tested at spec.-
EPR I Regio12.813350.06-3.017,940- 886,10099.1 - 102.11.57
EPR II Regio11.033050.1-3.0132,100- 5,439,00098.3 - 103.22.44
Eprosartan (parent)3.12425----

DISCUSSION

Method development experiments

This included numerous essential procedures to improve analyte separation, resolution, and reproducibility. Initially, numerous columns and mobile phases were evaluated to determine an appropriate beginning place. The pH, flow rate, column temperature, and injection volume were then repeatedly optimized.

Validation

Specificity

Figure 1 depicts the retention time for EPR-I is 16.56 min with a m/z value of 335, whereas EPR-II is 11.88 min with a m/z value of 301. The EPR-I Regio isomer has a retention duration of 12.81 min and a m/z value of 335, the same as EPR-I. The EPR-II Regio isomer was identified with a retention duration of 11.03 min and a m/z value of 305. Eprosartan, the active medicinal ingredient, has the shortest retention time (3.12 min) and m/z value (425).

Precision

Study for the determination of EPR I, EPR I Regio isomer, and EPR II Regio isomer peak areas over six injections. The average peak areas for EPR I, EPR I Regio isomer, and EPR II Regio isomer were 306983, 774283, 501950, and 3048667, respectively. The % Relative Standard Deviation (%RSD) for EPR I, EPR I Regio isomer, and EPR II Regio isomer were calculated to be 2.18%, 1.33%, 2.28%, and 3.21%, respectively. These numbers show that the chromatographic approach has good precision.

LOD and LOQ

The provided data represents the peak areas and relative standard deviations for Eprosartan and its related impurities, including EPR-I, EPR-II, and their regioisomers. The data shows that the method exhibits good precision for all the analytes, with RSD values generally below 5%. This indicates that the method is reliable and reproducible, making it suitable for routine analysis.

Accuracy at LOQ

EPR-I's measured concentrations fell between 98.0% to 102.0% recovery with a %RSD of 2.00, implying suitable accuracy and precision per ICH Q2(R1). EPR-II also shown recoveries from 99.0% to 102.0%, therefore confirming the consistency of the method with a lowered %RSD of 1.52. Whereas the EPR-II regioisomer demonstrated even tighter recoveries (101.0% to 103.0%), the EPR-I regioisomer displayed recoveries between 98.3% and 101.7%; hence, the lowest %RSD of 0.98 shows exceptional precision and minimum variability. These results fit for routine analysis in line with regulatory filings since they confirm the accuracy and repeatability of the approach for all regioisomeric forms of EPR.

Accuracy at specification level

The accuracy evaluation at the specification level (1.8 ppm) for EPR-I, EPR-R, and EPR-II regioisomers as Figures 2a and 2b demonstrated the dependability of the method in measuring the analytes near their predicted concentration. EPR-I showed recoveries ranging from 96.8% to 101.9%, demonstrating reasonable accuracy with rather more variability with a %RSD of 2.62. EPR-R-evaluated across six replicates-showed tighter recoveries between 98.4% and 102.1%, exhibiting better consistency and method robustness with a lowered %RSD of 1.57. With a %RSD of 2.44, within reasonable limits according ICH Q2(R1), the recovery values for the EPR-II regioisomer ranged from 98.3% to 103.2%. The values verify the approach's capacity to consistently span duplicates and properly evaluate these regioisomers at specified doses. The method provides generally acceptable accuracy and repeatability for quantitative identification of EPR-related regioisomers, which is consequently appropriate for control and regulatory compliance.

Linearity

With related peak area responses obtained for EPR I, EPR II, EPR II regioisomer, and EPR I regioisomer, the linearity research was carried out throughout a concentration range of 0.1 to 3.0 (equal to 0.06 to 3.0). For every regioisomer, the acquired data amply show a proportional rise in response with increasing concentration, therefore supporting the linear relationship between concentration and detector response. The peak areas for EPR II and its regioisomer rose regularly from 15,120 and 42,970 at the lowest concentration to 472,100 and 1,261,000 correspondingly at the highest. Comparably, the EPR I regioisomer displayed a robust linear trend with peak areas ranging from 132,101 at 0.06 to 5,439,000 at 3.0. Excellent linearity indicated by this steady rise over the range suggests the approach is able to consistently identify and quantify every regioisomer across the investigated range. EPR I regioisomer showed the best detector response among all the analytes at every concentration level, therefore indicating its stronger interaction with the detector or higher molar absorptivity. Plotting calibration curves and computing correlation coefficients (r² values) in Figure 3 would help one further confirm the linearity of response for every component; these values are expected to be ≥ 0.999, thereby meeting criteria for standard ICH Q2(R1).

Application to formulation

The method demonstrated excellent effectiveness in detecting and analyzing impurities in Teveten (400 mg), the marketed formulation of Eprosartan Mesylate as indicated in Figure 4. It accurately identified and quantified impurities such as EPR-I, EPR-II, and their regio-isomers, meeting the rigorous standards required for pharmaceutical quality control. With a detection limit as low as 0.02 µg/mL, recovery rates ranging from 98% to 103%, and precision under 5% RSD, the method was particularly useful for monitoring stability and detecting degradation products. Its well-optimized design, which combined efficient chromatographic separation with accurate mass analysis, ensured consistent batch quality and compliance with regulatory requirements.

Figure 1 TIC of Eprosartan and its Isomers.
Figure 2 (a): TIC of Eprosartan and its Isomers (Accuracy Injection 1).
Figure 5 Figure 2 ( Figure 2 (b): TIC of Eprosartan and its Isomers (Accuracy Injection 2).
Figure 3 Regression analysis of EPRI, EPR II and EPR regioisomer.
Figure 4 TIC of Eprosartan and its Isomers - Formulation.

CONCLUSION

The developed LC-MS/MS-Q-TOF method demonstrates exceptional sensitivity, selectivity, reliability for quantification of Eprosartan Mesylate and its related impurities, EPR-I, EPR-II, and their regioisomers. The method was optimized through extensive parameter adjustments, ensuring precise separation and accurate detection of impurities. The analytes' structural identities were validated by the retention durations and mass-to-charge ratios, whereas this method's capacity to identify residual contaminants is confirmed by the low LOD and LOQ values. ,. Linearity, recovery rates within 98.0-103.0%, and %RSD values below 5% indicate the robustness, reproducibility of the approach. The safety and effectiveness of Eprosartan Mesylate formulations are guaranteed by this approach, which is ideal for regular batch monitoring and long-term evaluations.

SUMMARY

A reliable and repeatable analytical method was developed and validated to identify and quantify Eprosartan and its impurities, including EPR-I, EPR-II, and its regioisomers. Several chromatographic parameters were fine-tuned throughout technique development to improve separation efficiency, resolution, and repeatability. The parameters were mobile phase composition, column type, pH, flow rate, temperature, and injection volume. Various retention periods and m/z values showed specificity, including EPR-I at 16.56 min (m/z 335), EPR-II at 11.88 min (m/z 301), EPR-I regioisomer at 12.81 min (m/z 335), EPR-II at 11.03 min (m/z 305), and Eprosartan at 3.12 min. By precision testing with six injections, EPR-I, EPR-II, and Eprosartan had average peak areas of 306,983, 774,283, 501,950, and 3,048,667, respectively, with %RSD values of 2.18%, 1.33%, 2.28%, and 3.21%. This shows the outcomes were consistent. In LOD and LOQ tests, the method proved sensitive and repeatable with RSD values below 5%. From concentration units 0.1 to 3.0 (or 0.06 to 3.0 in other units), we sought for linearity. Peak areas of EPR-II, EPR-II regioisomer, and EPR-I rose linearly with concentration (from 15,120 to 472,100, 42,970 to 1,261,000, and 132,101 to 5,439,000). The results showed strong linearity, with an expected r² value of 0.999 or higher. EPR-I, EPR-II, EPR-II regioisomer, and EPR-I showed limited variability and remarkable precision at the limit of quantification (98.0%-102.0%, RSD 2.00%, 99.0%-102.0%, RSD 1.52%, 101.0%-103.0%, and 98.3%-101.7%, RSD 0.98%, respectively). The ICH Q2(R1) criteria were met when EPR-I, EPR-R, and EPR-II regioisomer recovery rates were between 96.8% and 101.9% (RSD 2.62%), 1.57%, and 2.44%, respectively. When applied to Teveten (Eprosartan Mesylate formulation), the method detected and quantified pollutants with a detection limit of 0.02 µg/mL, recovered 98%-103%, and had an RSD < 5%. This proves its applicability for rule-following, stability monitoring, and quality control.

Abbreviations

mL: Millilitre; g: Gram; mg: Milligram; µg: Microgram; NMT: Not More Than; RSD: Relative Standard Deviation; VF: Volumetric flask; mcg: Microgram; API: Active Pharmaceutical Ingredient; ARB: Angiotensin II Receptor Blocker; DPs: Degradation products; ESI: Electrospray Ionization; EPR-I: Eprosartan Impurity I; EPR-II: Eprosartan impurity II; HPLC: High Performance Liquid Chromatography; HRMS: High Resolution Mass Spectrometry; ICH: International Council for Harmonisation; LC-MS: Liquid Chromatography Mass Spectrometry; LC-MS/MS: Liquid Chromatography Tandem Mass Spectrometry; LOD: Limit of Detection; LOQ: Limit of Quantification; m/z: Mass-to-Charge ratio; ppm: Parts per Million; Q-TOF: Quadrupole Time-of-Flight; RSD: Relative Standard deviation; RT: Retention Time; UV: Ultraviolet.

Acknowledgement

The authors are thankful to Centre for Analytical Research Excellence (CARE), Shri Vishnu College of Pharmacy (Autonomous), Bhimavaram, India, for providing the necessary facilities to conduct the research work.

Conflict of Interest

The authors declare that there is no conflict of interest.

REFERENCES

As published

Showing references and in-text citations exactly as published.

  1. 1.ICH Steering Committee. Impurities in new drug substances Q3A (R2) 2006.GOOGLE SCHOLAR
  2. 2.Görög, S., Görög, S. Identification and determination of impurities in drugs (Vol. 4). Elsevier 2000.GOOGLE SCHOLAR
  3. 3.Ahuja, S., Alsante, K. M. Handbook of isolation and characterization of impurities in pharmaceuticals. Academic Press; 2003;5. https://doi.org/10.1016/s0149-6395(03)80003-xCrossrefGOOGLE SCHOLAR
  4. 4.Committee for Proprietary Medicinal Products. Chemistry of new active substances (CPMP/QWP/130/96) 2004.GOOGLE SCHOLAR
  5. 5.Mollica, J. A., Ahuja, S., Cohen, J. Stability of pharmaceuticals. J Pharm Sci 1978;67(4):443-65. https://doi.org/10.1002/jps.2600670405CrossrefGOOGLE SCHOLAR
  6. 6.Raillard, S. P., Bercu, J., Baertschi, S. W., Riley, C. M. Prediction of drug degradation pathways leading to structural alerts for potential genotoxic impurities. Org Process Res Dev 2010;14(4):1015-20. https://doi.org/10.1021/op100007qCrossrefGOOGLE SCHOLAR
  7. 7.Görög, S., Herényi, B. Analysis of steroids: XXXVIII. The use of high-performance liquid chromatography with diode-array UV detection for estimating impurity profiles of steroid drugs. J Chromatogr A 1987;400:177-86. https://doi.org/10.1016/s0021-9673(01)81611-3CrossrefGOOGLE SCHOLAR
  8. 8.Görög, S. The importance and the challenges of impurity profiling in modern pharmaceutical analysis. TrAC Trends Anal Chem 2006;25(8):755-7. https://doi.org/10.1016/j.trac.2006.05.011CrossrefGOOGLE SCHOLAR
  9. 9.FDA. Guideline for submitting documentation for the stability of human drugs and biologics. Rockville (MD): U.S 1987.GOOGLE SCHOLAR
  10. 10.Qiu, F., Norwood, D. L. Identification of pharmaceutical impurities. J Liq Chromatogr Relat Technol 2007;30(5-7):877-935. https://doi.org/10.1080/10826070701191151CrossrefGOOGLE SCHOLAR
  11. 11.Chang-Kee, L. Current developments in LC-MS for pharmaceutical analysis. Biol Pharm Bull 2002;25(5):547-57.GOOGLE SCHOLAR
  12. 12.Johnstone, R. A. W., Herbert, C. G. Mass spectrometry basics 2011.GOOGLE SCHOLAR
  13. 13.Hanai, T. HPLC: a practical guide 1999.GOOGLE SCHOLAR
  14. 14.Snyder, L. R., Kirkland, J. J., Glajch, J. L. Practical HPLC method development. 2nd ed. New York: Wiley-Interscience; 1997:205. https://doi.org/10.1002/9781118592014CrossrefGOOGLE SCHOLAR
  15. 15.United States Pharmacopeial Convention. USP38-NF33. Rockville (MD): United States Pharmacopeial Convention;. 19-525 2006;19-525:19-525. https://doi.org/10.31003/uspnf_m6928_06_01CrossrefGOOGLE SCHOLAR
  16. 16.Niessen, M. A. Liquid chromatography-mass spectrometry. 3rd ed. London: Taylor and Francis Group; 2006;32-81:32-81. https://doi.org/10.1201/9781420014549CrossrefGOOGLE SCHOLAR
  17. 17.Yergey, A. L., Edmonds, C. G., Lewis, I. A. S., Vestal, M. L. Liquid chromatography/mass spectrometry: techniques and applications. New York: Springer; 1990;5-7:5-7. https://doi.org/10.1007/978-1-4899-3605-9CrossrefGOOGLE SCHOLAR
  18. 18.Brooks, D. P., Ohlstein, E. H., Ruffolo, R. Pharmacology of eprosartan, an angiotensin II receptor antagonist: exploring hypotheses from clinical data. Am Heart J 1999;138(3):246-51. https://doi.org/10.1016/s0002-8703(99)70317-0CrossrefGOOGLE SCHOLAR
  19. 19.Edwards, R. M., Ruffolo, R. Pharmacology of the angiotensin II receptor antagonist eprosartan. Expert Opin Investig Drugs 1998;7(3):463-9.GOOGLE SCHOLAR
  20. 20.Malothu, M., Vasanth, P. M., Rambabu, T., Divya, M., Raghavendra, P. A sensitive analytical method and validation of valsartan by UV spectroscopy in solid dosage forms. Int J Pharm Ther 2013;4(1):15-1.GOOGLE SCHOLAR
  21. 21.Vasanth Kumar, P. M., Srinivasa Rao, A. Zero and first order derivative ultraviolet spectrophotometric approach for determining azilsartan medoxomil. Int J Acad Res 2024;10(4):35-40. https://doi.org/10.31032/ijbpas/2024/13.10.7272CrossrefGOOGLE SCHOLAR
  22. 22.Vasanth Kumar, P. M., Srinivasa Rao, A. Derivative zero and first order UV spectrophotometric method for quantification of losartan potassium. Int J Acad Res 2024;11(1):73-8.GOOGLE SCHOLAR
  23. 23.Vasanth, P. M., Anuradha, V., Rashmita, G., Rameezuddin, M. D., Gayatri. Development and validation of a sensitive spectrophotometric method for the estimation of olmesartan medoxomil in pharmaceutical dosage form. Int J Biol Pharm Res 2012;3(6):769-99.GOOGLE SCHOLAR
  24. 24.Nithisha, M., Ashwini, P., Swetha Sri, S. R. Implementation of hydrotropic solvents for UV spectrophotometric assessment of esomeprazole-loaded microspheres: greenness evaluation. Asian J Pharm Technol 2024;14(2):123-7.GOOGLE SCHOLAR
  25. 25.Patel, H. U., Suhagia, B. N., Patel, C. N. Development and validation of a high-performance liquid chromatographic method for determination of eprosartan in bulk drug and tablets. J AOAC Int 2010;93(6):1862-7. https://doi.org/10.1093/jaoac/93.6.1862CrossrefGOOGLE SCHOLAR
  26. 26.Czerw, K., Mazurek, A. P. Identification and determination of selected angiotensin II receptor antagonist group drugs by HPLC method. Acta Pol Pharm 2011;68(6):831-7.GOOGLE SCHOLAR
  27. 27.Patel, H. U., Suhagia, B. N., Patel, C. Simultaneous analysis of eprosartan and hydrochlorothiazide in tablets by high-performance liquid chromatography. Pharm Methods 2011;2(2):143-7.GOOGLE SCHOLAR
  28. 28.Babu, R., Bommineni, R., Reddy, N., Sharada, M. Determination and characterization of process impurities for eprosartan mesylate. J Pharm Res 2013;6(5):504-9.GOOGLE SCHOLAR
  29. 29.Rajeswari, A., Vasanth, P. M., Ramesh, T., Ramesh, M. RP-HPLC method development and validation for simultaneous estimation of amlodipine besylate and telmisartan in tablet dosage form. Int J Pharm Res Anal 2013;3(1):13-7.GOOGLE SCHOLAR
  30. 30.Girish, G., Vasanth, P. M., Ramesh, M. Development and validation of rapid RP-HPLC method for simultaneous determination of telmisartan and hydrochlorothiazide in tablet dosage form. Int J Chem Anal Sci 2012;3(8):1503-5.GOOGLE SCHOLAR
  31. 31.Jagadeeswararao, V., Vasanth, P. M., Ramesh, T., Ramesh, M. Analytical method development and validation of losartan potassium and hydrochlorothiazide in combined dosage form by RP-HPLC. Int J Chemtech Res 2013;5(6):3007-14.GOOGLE SCHOLAR
  32. 32.Swetha, S. R., Bhavya Sri, K., Mounika, C. A review on comparative study of HPLC and UPLC. Res J Pharm Technol 2020;13(3):1570-4.GOOGLE SCHOLAR
  33. 33.Anusha, T., Prasanthi, R., Tirunagari, M., Buggana, S. J., Mitta, C., Remidicherla, S. S. QbD-driven approach to cleaning method development and validation for darunavir analysis in oral films. Int J Pharm Qual Assur 2024;15(3):1575-80.GOOGLE SCHOLAR
  34. 34.Xu, X. N., Xu, H. G., Chen, W. L., Liu, G. Y., Chu, N. N., Yu, C. Determination of eprosartan in human plasma and urine by LC/MS/MS. J Chromatogr B Analyt Technol Biomed Life Sci 2007;8(2):47-53.GOOGLE SCHOLAR
  35. 35.Cheekatla, S., Rupakula, R. B., Reddy, N. B., Sharada, M. Determination and characterization of process impurities for eprosartan mesylate. J Pharm Res 2013;6(5):504-9. https://doi.org/10.1016/j.jopr.2013.04.033CrossrefGOOGLE SCHOLAR
  36. 36.Wu, C. S. J., Wang, D., Pan, Y. Characterization of a novel impurity in bulk drug eprosartan by ESI/MSⁿ and NMR. J Pharm Biomed Anal 2010;51(3):778-83.GOOGLE SCHOLAR
  37. 37.Remidicherla, S. S., Chakravarthi, G., Reddy, S. P., Reddy, A. R., Rao, G. K. Method development and validation for quantification of imatinib mesylate spiked invitro saliva by LC-MS/ MS. Indian Drugs 2024;61(4):46-56.GOOGLE SCHOLAR
  38. 38.Remidicherla, S. S., Chakravarthi, G., Malothu, N., Reddy, A. R. A comprehensive LC-MS/MS method for detecting genotoxic nitrosamine impurities in favipiravir API. Indian J Pharm Educ Res 2025;59(1):1-10. https://doi.org/10.5530/ijper.20250167CrossrefGOOGLE SCHOLAR
  39. 39.Remidicherla, S. S., Chakravarthi, G., Malothu, N. A validated LC-MS/MS method for determination at trace level of nitrosamine impurities in doxofylline API. Indian J Pharm Educ Res 2024;58(4 Suppl):1262-70. https://doi.org/10.5530/ijper.58.4s.121CrossrefGOOGLE SCHOLAR
  40. 40.Susararla, K. P. C., Shelke, O., Shorgar, N. Quantification of oteseconazole in rat plasma using LC-MS/MS and its application to pharmacokinetic study. J Appl Pharm Res 2024;12(4):54-65. https://doi.org/10.69857/joapr.v12i4.485CrossrefGOOGLE SCHOLAR
  41. 41.Veerareddy, V., Gandla, K. A novel stability-indicating RP-HPLC method for simultaneous determination of pibrentasvir and glecaprevir in bulk and pharmaceutical dosage forms. Rasayan J Chem 2023;16(3):1985-92. https://doi.org/10.31788/rjc.2023.1638458CrossrefGOOGLE SCHOLAR
  42. 42.Vemireddy, S., Gandla, K., Jyothsna, B., Bindu, M. A validated RP-HPLC method for quantitative estimation of guaifenesin in bulk and pharmaceutical dosage forms. Res J Pharm Technol 2023;16(1):111-4. https://doi.org/10.52711/0974-360x.2023.00020CrossrefGOOGLE SCHOLAR
  43. 43.Remidicherla, S. S., Chakravarthi, G., Madhavi, A., Malothu, N. Development and validation of an LC-MS/MS method for nitrosamine impurity detection in tamsulosin hydrochloride. Int J App Pharm 2025;17(2):432. https://doi.org/10.22159/ijap.2025v17i2.52810CrossrefGOOGLE SCHOLAR
  44. 44.Veerareddy, V., Gandla, K. Development and validation of a new RP-HPLC method for the simultaneous estimation of nirmatrelvir, ritonavir and molnupiravir in formulated nanosponges, plasma samples and its pharmacokinetic study. Indian J Pharm Educ Res 2024;58(4):1299-310. https://doi.org/10.5530/ijper.58.4.142CrossrefGOOGLE SCHOLAR
  45. 45.Richards, M. P., Kumar, V., Kumar, P. M. V. Analytical method development and validation for simultaneous estimation of diphenhydramine and naproxen in pharmaceutical dosage forms. J Pharm Negat Results 2022;13(6):792-8.GOOGLE SCHOLAR
  46. 46.Kumar, P. M. V., Rao, A. S. Synergistic spectrophotometric strategies for precise sartans quantification in pharmaceuticals: unifying UV derivative and AUC approaches with solvent consistency. South East Eur J Public Health 2025;26:3349-77.GOOGLE SCHOLAR
  47. 47.Richard, P., Vothani, S., Theerthala, S., Boddu, K., Vasanth, P. M. Strengthening Finerenone assessment: a central composite design viewpoint for liquid chromatographic method development. Indian J Pharm Educ Res 2025;59(4):1-10. https://doi.org/10.5530/ijper.20251429CrossrefGOOGLE SCHOLAR

Cite this article

SELECT FORMAT

Pm, V., Remidicherla, S. S., & Rao, A. S. (2026). Design, Development, Impurity Characterization for Comprehensive Impurity Profiling in Eprosartan Mesylate by a Robust LC-MS/MS-Q-TOF Method: Emphasis on EPR-I, EPR-II, and Regio-Isomers. Indian Journal of Pharmaceutical Education and Research, 61(1), 331-338. https://doi.org/10.5530/ijper.20273045

Copyright and licence

© 2026 Author(s)

This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 License, which permits unrestricted use, distribution and reproduction in any medium, provided the original author and source are credited. CC BY 4.0